Addressing Ventilator Alarms while Transporting Patients within the Hospital

Recently, a patient become disconnected from the ventilator during transport in the hospital. Though the ventilator alarmed, the transport team was not able to quickly discover the problem.

Any alarm associated with life support must be addressed in real time. With the need to wear PPE (face shields and masks) and challenges with hearing audible alarms, it is even more imperative that we visualize and react to alarms. Ventilator alarms during transport must be identified in full and corrected. All transporting members have the authority to STOP THE LINE to address alarms and assess the patient at any time during the transport. It is common for multiple alarms to alert simultaneously so identification of all active alarms is critical. 

To improve troubleshooting of the ventilator alarms:

 

Trilogy Ventilator Alarm Troubleshooting Tip with Grab N’ Go Oxygen Cylinder:

Low Oxygen Inlet Pressure alarm while using 100% FiO2 on high settings